Skip to main content

Table 3 Recovery of perchlorate in synthetic matrix solutions after preparation

From: Validation data for the determination of perchlorate in water using ion chromatography with suppressed conductivity detection

Sample

Matrix-free standard; direct injection

Matrix-free standarda

100 mg/L chloride/sulphatea

500 mg/L chloride/sulphatea

1.0 µg/L ClO 4

 Recb area (%)

100 ± 1.19

88.8 ± 0.28

83.8 ± 4.6

84.7 ± 4.4

 Recb height (%)

100 ± 0.89

91.3 ± 0.41

91.3 ± 3.5

88.9 ± 2.9

1.5 µg/L ClO 4

 Recb area (%)

100 ± 0.32

102 ± 4.9

86.0 ± 4.1

106 ± 4.5

 Recb height (%)

100 ± 0.49

101 ± 3.7

87.6 ± 1.9

110 ± 4.0

5.0 µg/L ClO 4

 Recb area (%)

100 ± 0.76

96.1 ± 1.9

95.7 ± 2.6

97.0 ± 2.0

 Recb height (%)

100 ± 0.76

98.8 ± 0.84

96.6 ± 0.82

98.7 ± 0.71

  1. apre-treated with three layer cartridge
  2. bRec: recovery of the perchlorate signal in sample [Rec Peak Area = 100 × Area(Sample)/Area(Standard); Rec Peak Height = 100 × Height(Sample)/Height(Standard)]