Skip to main content

Table 1 Comparison of performance data obtained applying various injection volumes

From: Validation data for the determination of perchlorate in water using ion chromatography with suppressed conductivity detection

Performance data 1 µg/L to 10 µg/L

Injection volume and signal evaluated

500 µL

750 µL

1000 µL

Area

Height

Area

Height

Area

Height

Linearity of the function

Yes

Yes

Yes

Yes

Yes

Yes

s x0 (µg/L)

0.128

0.106

0.105

0.171

0.0603

0.0659

V x0 (%)

2.33

1.93

1.90

3.12

1.10

1.20

LOD (µg/L)

0.512

0.424

0.420

0.684

0.241

0.264

LOQ (µg/L)

1.54

1.27

1.26

2.05

0.724

0.791

  1. s x0 standard deviation of the procedure [24]; V x0 Variation coefficient of the procedure (relative s x0) [24]; LOD approximate value for the limit of detection (LOD = 4·s x0, [25]); LOQ approximate value for the limit of quantification (LOD = 12·s x0, [25])